The Via Nano 6 requires special precautions regarding EMC and needs to be installed and put into service according to the EMC information provided in these accompanying documents.
WARNING: Portable RF communications equipment (including peripherals such as antenna cables and external antennas) should be used no closer than 30 cm (12 inches) to any part of the Via Nano 6, including cables specified by the manufacturer. Otherwise, degradation of the performance of this equipment could result.
WARNING: A risk of increased emissions or decreased immunity may result if any additional cables are attached.
WARNING: This device has not been tested for compatibility with all other potential RF Emitters such as X-ray, Metal Detectors, Electrosurgical Equipment, Diathermy Equipment, 5G Cellular, NFC, WPT, or Electronic Article Surveillance (EAS) devices. Caution should be used if such emitters are present within the use environment.
WARNING: Use of this equipment adjacent to or stacked with other equipment should be avoided because it could result in improper operation. If such use is necessary, this equipment and the other equipment should be observed to verify that they are operating normally.
WARNING: Use of accessories, transducers, and cables other than those specified or provided by the manufacturer of this equipment could result in increased electromagnetic emissions or decreased electromagnetic immunity of this equipment and result in improper operation.
Guidance and manufacturer’s declaration – electromagnetic emissions | ||
The Via Nano 6 is intended for use in the electromagnetic environment specified below. The customer or the user of the Via Nano 6 should ensure that it is used in such an environment. | ||
Emissions test | Compliance | Electromagnetic environment - guidance |
RF emissions CISPR 11 | Group 1 | The Via Nano 6 uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment. |
RF emissions CISPR 11 | Class B | The Via Nano 6 is suitable for use in all establishments, including domestic establishments and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes. |
Harmonic emissions IEC 61000-3-2 | Class A | |
Voltage Fluctuations IEC 61000-3-3 | Complies | |
Guidance and manufacturer’s declaration – electromagnetic immunity | |||
The Via Nano 6 is intended for use in the electromagnetic environment specified below. The customer or the user of the Via Nano 6 should ensure that it is used in such an environment. | |||
Immunity test | IEC 60601 test level | Compliance level | Electromagnetic environment - guidance |
Electrostatic discharge (ESD) IEC 61000-4-2 | +8kV contact +15kV air | +8kV contact +15kV air | Floors should be wood, concrete or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30%. |
Electrical fast transient/burst IEC 61000-4-4 | +2kV for power supply lines | +2kV for power supply lines | Mains power quality should be that of a typical commercial or hospital environment |
Surge IEC 61000-4-5 | +1kV line(s) to line (2) +2kV line(s) to earth | +1kV line(s) to line (2) | Mains power quality should be that of a typical commercial or hospital environment. |
Voltage dips, short interruptions and voltage variations on power supply input lines IEC 61000-4-11 | 0 % UT; 0,5 cycle at 0°, 45°, 90°, 135°, 180°, 225°, 270° and 315° 0 % UT; 1 cycle and 70 % UT; 25/30 cycles Single phase: at 0° 0 % UT; 250/300 cycle | 0 % UT; 0,5 cycle at 0°, 45°, 90°, 135°, 180°, 225°, 270° and 315° 0 % UT; 1 cycle and 70 % UT; 25/30 cycles Single phase: at 0° 0 % UT; 250/300 cycle | Mains power quality should be that of a typical commercial or hospital environment. If the user of the Via Nano 6 requires continued operation during power mains interruptions, it is recommended that the Via Nano 6 be powered from an uninterruptible power supply or a battery. |
Power frequency (50/60 Hz) magnetic field IEC 61000-4-8 | 30 A/m | 30 A/m | Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment. |
NOTE: UT is the A.C. mains voltage prior to application of the test level. | |||
Guidance and manufacturer’s declaration – electromagnetic immunity | |||
The Via Nano 6 is intended for use in the electromagnetic environment specified below. The customer or the user of the Via Nano 6 should ensure that it is used in such an environment. | |||
Immunity test | IEC 60601 test level | Compliance level | Electromagnetic environment - guidance |
Conducted RF IEC 61000-4-6 Radiated RF IEC 61000-4-3 | 3 Vrms 0,15 MHz – 80 MHz 6 Vrms in ISM and amateur radio bands between 0,15 MHz and 80 MHz 80 % AM at 1 kHz 10 V/m 80 MHz to 2.7 GHz | 3 Vrms 0,15 MHz – 80 MHz 6 Vrms in ISM and amateur radio bands between 0,15 MHz and 80 MHz 80 % AM at 1 kHz 10 V/m 80 MHz to 2.7 GHz | Portable and mobile RF communications equipment should be used no closer to any part of the Via Nano 6, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter. Recommended separation distance: d = [3.5/10] √P 80 MHz to 800 MHz d = [7/10] √P 800 MHz to 2.7 GHz where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in meters (m). Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey, a should be less than the compliance level in each frequency range. b Interference may occur in the vicinity of equipment marked with the following symbol: |
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies. NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people. | |||
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the Via Nano 6 is used exceeds the applicable RF compliance level above, the Via Nano 6 should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating the Via Nano 6. | |||
Guidance and manufacturer’s declaration – electromagnetic immunity | |||||||
The Via Nano 6 is intended for use in the electromagnetic environment specified below. The customer or the user of the Via Nano 6 should ensure that it is used in such an environment. | |||||||
Immunity test | IEC 60601 test level | Compliance level | Electromagnetic environment - guidance | ||||
IMMUNITY to proximity fields from RF wireless communications equipment IMMUNITY to proximity magnetic fields IEC 61000-4-39 | MHz | Modulation | Field Strength | MHz | Modulation | Field Strength | Portable and mobile RF communications equipment should be used no closer to any part of the Via Nano 6, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter. Recommended separation distance: E = [6/d] √P d = [6/E] √P where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer, d is the recommended separation distance in meters (m), and E is the field strength in V/m. Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey, should be less than the compliance level in each frequency range. Interference may occur in the vicinity of equipment marked with the following symbol: |
385 450 710 745 780 810 870 930 1720 1845 1970 2450 5240 5500 5785 0.1342 13.56 0.030 | 18 Hz 18 Hz 217 Hz 217 Hz 217 Hz 18 Hz 18 Hz 18 Hz 217 Hz 217 Hz 217 Hz 217 Hz 217 Hz 217 Hz 217 Hz 2.1kHz 50kHz CW | 27 V/m 28 V/m 9 V/m 9 V/m 9 V/m 28 V/m 28 V/m 28 V/m 28 V/m 28 V/m 28 V/m 28 V/m 9 V/m 9 V/m 9 V/m 65 A/m 7.5 A/m 8 A/m | 385 450 710 745 780 810 870 930 1720 1845 1970 2450 5240 5500 5785 0.1342 13.56 0.030 | 18 Hz 18 Hz 217 Hz 217 Hz 217 Hz 18 Hz 18 Hz 18 Hz 217 Hz 217 Hz 217 Hz 217 Hz 217 Hz 217 Hz 217 Hz 2.1kHz 50kHz CW | 27 V/m 28 V/m 9 V/m 9 V/m 9 V/m 28 V/m 28 V/m 28 V/m 28 V/m 28 V/m 28 V/m 28 V/m 9 V/m 9 V/m 9 V/m 65 A/m 7.5 A/m 8 A/m | ||
NOTE: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people. | |||||||
Recommended separation distances between portable and mobile RF communications equipment as well as RF wireless communications equipment and the Via Nano 6. | ||||
The Via Nano 6 is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the Via Nano 6 can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the Via Nano 6, as recommended below, according to the maximum output power of the communications equipment. | ||||
Rated maximum output power of transmitter (W) | Separation distance according to frequency of transmitter (m) | |||
80 to 800 MHz d = [3.5/10] √P | 800 MHz to 2.7 GHz d = [7/10] √P | 710, 745, 780, 5240, 5500, 5785 MHz d = [6/9] √P | 385, 450,810, 870, 930, 1720, 1845, 1970, 2450 MHz d = [6/28] √P | |
0.01 | 0.035 | 0.070 | 0.067 | 0.021 |
0.1 | 0.110 | 0.221 | 0.211 | 0.070 |
1 | 0.350 | 0.700 | 0.667 | 0.214 |
10 | 1.107 | 2.213 | 2.108 | 0.700 |
100 | 3.500 | 7.000 | 6.670 | 2.143 |
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer. NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies. NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people. | ||||